RES010 – Firesafe Cannula Valve

$0.00

Description

Each

  • Bidirectional
  • Helps protect against oxygen tubing fires
  • A thermal fuse (a.k.a. a firebreak) designed to stop the flow of oxygen in the event that the downstream oxygen tubing is ignited
  • Helps keep patients safe no matter what method of oxygen delivery they use
  • Offers a cost-effective solution to lower the risk of serious patient injury in an oxygen fire

Gas Compatibility Oxygen and oxygen enriched air
Maximum Operating Pressure 600 kPa*

*Maximum static pressure delivered by the gas supply source

Flow Rate 0.25* – 25 l/min

The Firesafe Cannula Valve can be safely used at lower flow rates but may not meet the internal leak rate of 10 ml/min when activated in the event of a fire.

Resistance to Flow ≤ 0.07 kPa at 2 l/min
≤ 0.3 kPa at 5 l/min
≤ 1 kPa at 10 l/min
Intended Life 5 years, expiry date engraved on device
Application Single patient multiple use
Packaging 100 units per box (5 bags per box, 20 units per bag)
Connections EN 13544-2, 6 mm nominal tubing connection
Environmental Limits for Transport and Storage Temperature: -25 °C to 70 °C (-4 °F to 140 °F) Humidity: 0 – 100 % RH
Environmental Limits for Use Temperature: 0 °C to 50 °C (32 °F to 122 °F) Humidity: 0 – 100 % RH
Applied Standards EN 13544-2:2002 – Respiratory Therapy Equipment – Part 2: Tubing and connectors.
Regulatory EC: MDD Class IIa Medical Device

USA: FDA Registered Class 1 Medical Device; 510(k) exempt

GMDN Code and Term: 60391 – Fire safety valve